Application method of negative plate of high frequency electric knife
1
The negative plate should be located at the muscle with abundant blood vessels close to the operation site (it is recommended to be located at the upper arm and thigh of the artery)
2
It is best to scrape off the hair first, then remove the oil and dander on the skin with alcohol, and then paste the circuit negative plate after drying to reduce the impedance value and achieve a safer effect
3
When the surgical site of the patient is determined, select the position closest to the surgical site without hindering the operation, avoid touching the conductive adhesive surface with fingers, pull it open and carefully press the four corners
4
Use the circuit board with wire: insert the connector of the circuit board correctly into the socket seat of the electric knife circuit
When using the circuit board without wire, check whether the return line is good
Open the collet and place the circuit patch to the deep part of the conductive adhesive, and then clamp it, and then pull it gently to ensure that it will not loosen
Insert the connector into the return socket
5
Paste the circuit board vertically along the longitudinal position of the blood vessel distribution as far as possible, which is conducive to the recovery of current
6
If the circuit board is pasted on the upper arm or thigh, avoid the circuit board that will not touch or overlap itself
7
Before the skin electrode is ready to be placed, check whether the electrode and wire are damaged or expired, so as to avoid accidents
8
The negative plate shall not be attached to the protruding parts of bones, physiological bends, dry skin, joints and artificial implanted metal parts or close to the electrode line of ECG, so as to avoid fever and scald caused by poor current circuit
9
The negative plate shall not be attached to the waist, armpit and compressed parts of the body, which is not conducive to heat dissipation and will cause thermal burns due to heat accumulation
10
The corners of the negative electrode plate shall not be rolled up
The reduction of the contact area between the electrode plate and the human body will increase the impedance during current recovery, which will lead to frequent alarm of the electric knife due to excessive impedance, no output of the knife head or burn
11
The pasting position of the negative plate shall not be in line with the operation position through the chest, especially for patients with heart disease and patients with stents
premiumexporter.com温馨提示
本网页所展示的有关【Disposable medical high frequency electric knife skin electrode negative plate strip wire】的信息/图片/参数等由的会员【Wuhan Mailang Medical Technology Co., Ltd】提供,由premiumexporter.com会员【Wuhan Mailang Medical Technology Co., Ltd
】自行对信息/图片/参数等的真实性、准确性和合法性负责,本平台(本网站)仅提供展示服务,请谨慎交易,因交易而产生的法律关系及法律纠纷由您自行协商解决,本平台(本网站)对此不承担任何责任。您在本网页可以浏览【Disposable medical high frequency electric knife skin electrode negative plate strip wire】有关的信息/图片/价格等及提供【Disposable medical high frequency electric knife skin electrode negative plate strip wire】的商家公司简介、联系方式等信息。