The new ENVITEC anvit oxygen battery oom202 is applicable to pb840 / pb760, Raphael machine, etc
The new aluminum box packed oxygen battery oom202 is compatible with oxygen batteries m-04, go-04, r-36med, mox-9, psr-11-117-m, max-12, 6051-0000-222, etc
the measurement range of oxygen sensor oom202 (O2 sensor, oxygen battery, oxygen probe, oxygen electrode) of ENVITEC anvit oxygen battery in Germany: , Zero - 100% output signal in air: , 13-16mv circuit interface 3pin (Molex) accuracy and repeatability: , & lt; 1% volume oxygen, linear error after calibration in 100% oxygen: , & lt; 3% relative error response time: , & lt; 12 seconds, zero drift relative to 90% relative value: , & lt; 200μ V effect of temperature in 100% ammonia for 5 minutes: , - Effect of 0
03% relative temperature and pressure: , Effect of mechanical shock proportional to oxygen partial pressure: , & lt; 1% relative error, working temperature after falling from a height of 1m: , 0 ℃ ~ 50 ℃ temperature compensation: , Built in thermistor to compensate for temperature influence: , 25 ℃ ~ 40 ℃, 3% relative error compensation (steady state): , 0 ℃ ~ 50 ℃, 8% relative error, working humidity: , Storage capacity of 0-99% RH (condensable air for several hours): , - It is recommended to use storage temperature at 20 ℃ ~ 50 ℃: , 5 ℃ ~ 15 ℃ recommended load: , & ge; 10Ω Preheating time: , & lt; 30 minutes, rated sensor life after sensor replacement: , & ge; 1000000% volume oxygen hourly weight: , About 28g Applicable to pb760 / pb840 machine, avea machine, Tianma machine, Hamerton machine and Aladdin machine In brief, the principle of oxygen battery is that the battery in which oxygen participates in redox will have the following electrode reaction: positive electrode: O2 + 2H2O + 4E - = 4OH - negative electrode: fe-3e - + 3OH - = Fe (OH) 3 general equation: 3O2 + 4Fe + 6H2O = 4Fe (OH) 3 when the measured oxygen concentration value deviates greatly from the set oxygen concentration value, the machine will give an alarm prompt, and then it can be calibrated, If the deviation is still large, the oxygen battery is generally exhausted and needs to be replaced
Generally, the service life of the oxygen battery is about 1 ~ 3 years
The oxygen battery is generally located between the patient's inspiratory end and the air oxygen mixer
For example, the air circuit system of MAQUET servo I machine: air and oxygen enter the two gas modules respectively, and are sent to the mixing chamber after being calculated according to the set moisture volume, pressure level and oxygen concentration
There is a mixing valve in the mixing chamber to stir the air flow
The air and oxygen are fully mixed and sent out after the actual oxygen concentration is monitored by the oxygen battery
Oxygen battery is only used for terminal monitoring in the machine, which has nothing to do with oxygen concentration control
2
Oxygen battery performance test (1) judge the failure of oxygen battery and observe the oxygen concentration monitoring
If the measured value deviates greatly from the set value, first check the oxygen source to see whether the oxygen pressure (0
24 ~ 0
6MPa, generally about 0
4MPa), flow (generally 120L / min) and oxygen concentration are normal
In addition, consider whether the air oxygen mixer is normal and eliminate the above possibilities, If the deviation between the measured value of oxygen concentration and the set value exceeds 10% for at least 30s, it can be judged that the oxygen battery has failed